RF Input Frequency
Final Procedure

Calibration Procedure For Test Set, Transponder Set An/Apm-305(A) (Ts-3395(A)) Page Navigation
  24    25    26    27    28  29  30  
img
. . . . .
TB 9-6625-2089-24
Figure 10. RF input frequency - equipment setup.
Table 10. RF Input Frequency Accuracy
Test
Synthesized signal generator
instrument
(MHz)
markers
Min
Max
2d
1084.8
1085.2
3d
1086.8
1087.2
4th
1088.8
1089.2
5th
1089.8
1090.2
6th
1090.8
1091.2
7th
1092.8
1093.2
8th
1094.8
1095.2
9th
1109.4
1110.6
22. Power Supply
a. Performance Check
(1) Connect multimeter between PS1A1TP1 (fig. 1) and ground. If multimeter does
not indicate between 27.75 and 28.25 V dc, perform b (1) below.
(2) Adjust autotransformer (A1) for 105 V ac.
Multimeter will indicate between
27.75 and 28.25 V dc.
(3) Adjust autotransformer for 125 V ac. Multimeter will indicate between 27.75