TB 9-6625-2089-24
Figure 9. RF input high power - equipment setup.
(21) Adjust MEASUREMENT DEMOD VID LEVEL control for MEASUREMENT,
METER indication of 33 dBW.
(22) Set variable attenuator No. 2 dial to setting on calibration chart for 43 dB at
1090 MHz.
(24) Adjust peak power meter to set reference to top of pulse as viewed on CH2 of
oscilloscope. If peak power meter does not indicate 0 1 dBm, perform b (11) through (14) below.
(25) Disconnect cable from MAIN RF IN/OUT and connect to AUX RF IN/OUT, set
DEMOD switch to DEMOD AUX and repeat (16) through (24) above for AUX RF IN/OUT.
b. Adjustments
(1) Set variable attenuator No. 2 dial to setting on calibration chart for 30 dB at
1090 MHz.
(2) Adjust signal generator output for a -2 dBm reference on peak power meter as
indicated on oscilloscope CH2.
(3) Adjust variable attenuator No. 2 dial to setting on calibration chart for 60 dB at
1090 MHz.
(4) Connect equipment as shown in figure 8, CONNECTION A.
27