TB 9-6625-2097-35
resources, annotate and affix DA Form 2417 (U. S. Army Calibration System Rejected
Instrument) and inform the owner/user accordingly in accordance with TB 750-25.
SECTION V
MODEL 13-4618-00
16. Preliminary Instructions
a. The
instructions outlined in
paragraphs 16 and
17 are preparatory to the
calibration
process. Personnel should become familiar with the applicable sections before beginning
the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see
NOTE
Unless otherwise specified, verify the, result of each test and,
whenever the test requirement is not met, take corrective
action before continuing with the calibration. Adjustments
required to calibrate the TI are included in this procedure.
Additional maintenance information is contained in the
manufacturer's manual for this TI.
NOTE
Unless otherwise specified, all controls and control settings
refer to the TI.
17. Equipment Setup
a. Remove protective cover from TI.
b. Slide all manual ink valves with tubing attached to ON position on recorder (A6).
c. Connect TI to recorder, using extender (B3).
e. Energize equipment and allow 30 minutes for warmup and stabilization.
12