SECTION I. IDENTIFICATION AND DESCRIPTION
SECTION III. CALIBRATION PROCESS FOR ME-30F/U

Calibration Procedure For Ac Voltmeter, Me-30F/U And Models 400F And 400Fl Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . . . .
TB 9-6625-2107-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
ME-30F/U - continued
Frequency response
(1 mV range)
10 Hz to 40 Hz
40 Hz to 500 kHz
500 kHz to 4 MHz
5%
1%
5%
Model 400F
Power input requirements
115 or 230 V ac, 50 to 400 Hz
100 V to 300 V
Ac voltage ranges (14)
Accuracy
(% FS + % reading)
(300 V to 300 V ranges)
Frequency response
20-40Hz  40-100Hz 100Hz-1MHz
1-2MHz
2-4MHz
(2 + 2)
(1 + 1)
(0.5 + 0.5)
(1 + 1)
(2 + 2)
(100 V range)
Frequency response
30 Hz 60 Hz
60 Hz 100 kHz
100 kHz 500 kHz
(2 + 2)
(1 + 1)
(1 +(+0,-7)
Model 400FL
Power input requirements
115 or 230 V ac, 50 to 400 Hz
100 V to 300 V
Ac voltage ranges (14)
% reading
Accuracy:
(300 V to 300 V ranges)
Frequency response
20-40Hz  40-100Hz 100Hz-1MHz
1-2MHz
2-4MHz
4
2
1
2
4
(100 V range)
Frequency response
30 Hz 60 Hz
60 Hz 100 kHz
100 kHz 500 kHz
4
2
+1, -8
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.  Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure.