TB 9-6625-2112-24
Figure 3. Amplitude - equipment setup.
(2) Set oscilloscope to view both input signals.
(3) Position controls as listed in (a) through (f) below:
(a) WIDTH switch to 1 s and vernier control fully ccw.
(b) REPETITION RATE switch to 10 kHz and vernier control fully cw.
(c) DELAY/ADVANCE switch to 10 s and vernier control fully cw.
(d) AMPLITUDE VOLTS switch to 10 V and vernier control fully cw.
(e) PULSE MODE switch to DELAY.
(f) PULSE POLARITY switch to POS.
(4) Measure pulse advance, using standard measurement techniques. If leading
edge of pulse output indicated on oscilloscope is not 10 s or greater to the right of sync
pulse leading edge, perform b (1) below.
(5) Position controls as listed in (a) and (b) below:
(a) REPETITION RATE switch to 1 kHz.
(b) DELAY/ADVANCE switch to 100 s and vernier control fully ccw.
(6) Measure pulse delay, using standard measurement techniques. If pulse delay
indicated on oscilloscope is not 10 s or less, perform b (2) below.
(7) Measure pulse delay/advance, using settings listed in table 7.
Oscilloscope
indications will be within limits specified.