TB 9-6625-2114-24
(4) Separate DCA OHMS and COM leads.
(5) Repeat (2) through (4) above until no further adjustments are necessary.
(6) Connect DCA OHMS and COM leads to resistance standard No. 1.
(7) Set RANGE switch to RX10.
(8) Adjust resistance standard for 1
indication on TI. Resistance standard will
indicate between 9.5 and 10.5 .
(9) Repeat technique of (7) and (8) above using settings listed in table 8. Resistance
standard indications will be within limits specified.
Test instrument
Resistance standard indications
RANGE
Indications
switch settings
(OHMS scale)
Min
Max
RX100
1
95
105
RX1K
1
950
1050
R10K
1
9.5
k
10.5 k
RX100K
1
95
k
105
k
RX1M
1
0.95
M
1.05 M
RX10M1
1
9.5
M
10.5 M
1Connect resistance standard no. 1 in series with resistance standard no. 2.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set FUNCTION switch to +DCA and RANGE switch to 150 MA.
(2) Connect DCA OHMS and COM leads to calibrator OUTPUT HI and LO.
(3) Adjust calibrator for a full scale indication on TI. Calibrator will indicate within
(4) Repeat technique of (3) above using settings and indications listed in table 9.
Test instrument
Calibrator indications
RANGE
switch settings
Indications
Min
Max
150
mA
150
145.5 mA
154.5 mA
50
mA
50
48.5 mA
51.5 mA
15
mA
15
14.55 mA
15.45 mA
5
mA
5
4.85 mA
5.15 mA
1.5
mA
1.5
1.455 mA
1.545 mA
0.5
mA
0.5
0.485 mA
0.515 mA
150
uA
150
145.5 uA
154.5 uA
50
uA
50
48.5 uA
51.5 uA
15
uA
15
14.55 uA
15.45 uA
5
uA
5
4.85 uA
5.15 uA
1.5
uA
1.5
1.455 uA
1.545 uA