TB 9-6625-2115-24
(15) Set OHMS FULL SCALE switch to 0.01.
(16) Place dc current shunt plugs in to 10 AMPERES position.
(17) Set ON-OFF switch to ON.
(18) Adjust ADJUST control for a 4.5 indication on TI OHMS meter 0 to 5 scale.
(19) Press and hold TEST pushbutton. TI OHMS meter will indicate between 4.25
and 4.75 on 0 to 5 scale. Release TEST pushbutton.
(20) Adjust ADJUST control fully ccw.
(21) Set ON-OFF switch to OFF.
(22) Connect equipment as shown in figure 2, connection B.
(23) Set OHMS FULL SCALE switch to 0.005.
(24) Place dc current shunt plugs in 100 AMPERES position.
(25) Set ON-OFF switch to ON.
(26) Adjust ADJUST control for a full-scale indication on TI meter 0 to 5 scale.
5