Figure 2. High resistance - equipment setup.

Calibration Procedure For Milliohmmeter, Model 673D Page Navigation
  1    2    3    4  5
TB 9-6625-2115-24
(27) Press and hold TEST pushbutton TI OHMS meter will indicate between 0.75
and 1.25 on 0 to 5 scale. Release TEST pushbutton.
b. Adjustments. No adjustments can be made.
9. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.