TB 9-6625-2119-35
(4) Set calibrator output to minimum.
b. Adjustments. Set calibrator for an output of 0.1900 V and adjust 190 mV DC CAL
ADJ R10 (fig. 1) until TI indicates 190.0 (R).
190 mV AC
CAL ADJ R25
200 V AC
FREQ
COMP
ADJ C52
2 V AC FREQ
COMP ADJ C50
20 V AC
FREQ
BATT TEST
COMP
ZERO ADJ R32
ADJ C51
190 mV DC CAL
ADJ R10
a. Performance Check
(1) Press AC/DC mode pushbutton to AC (in) and range 200 m pushbutton (in).
(2) Set calibrator for an output amplitude of 190 mV at an output frequency of 40 Hz.
If TI does not indicate within limits specified in first row of table 4, perform corresponding
(3) Repeat technique of (1) and (2) above, for calibrator outputs, TI settings and
indications listed in table 4. If TI does not indicate within limits specified in table 4,
perform corresponding adjustments in table 4.
Calibrator
Test instrument
Adjustments
Output
Output
Indication limits
amplitude
Range
Min
Max
190 mV
40 Hz
200 m
188.5
mV
191.5
mV
b (1) and (2)
190 mV
1 kHz
200 m
185.7
mV
194.3
mV
-----
190 mV
10 kHz
200 m
187.6
mV
192.4
mV
-----
1.9 V
40 Hz
2
1.885 V
1.915 V
-----
1.9 V
10 kHz
2
1.876 V
1.924 V
b (3) and (4)
-----
19 V
40 Hz
20
18.85 V
19.15 V