TB 9-6625-2119-35
(3) Set calibrator for a 2.2 V output.
(4) Position controls as listed in (a) through (c) below:
(a) POWER ON/OFF pushbutton pressed to ON.
(b) AC/DC mode pushbutton released to DC.
(c) A and V mode pushbuttons pressed to BAT (A and V pressed
simultaneously). If TI does not indicate 0.000 3 LSD, perform b (1) below.
(5) Disconnect calibrator and reinstall batteries. If TI display indicates negative,
perform b (2) below.
(6) Release POWER ON/OFF pushbutton to OFF.
b. Adjustments
(1) Adjust BATT TEST ZERO ADJ R32 (fig. 1) for a 0.000 indication on TI display.
(2) Install new batteries.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
8