(5) Press RECALL key and enter data 1 on keyboard. Adjust A4R13 (fig. 11) for
power meter indication of -5.0 dBm (R).
(6) Press RECALL key and enter data 4 on keyboard. Record power meter
indication.
(7) Press RECALL key and enter data 3 on keyboard. Adjust A4R9 (fig. 11) for
power meter indication of equal difference from +3 dBm as the value recorded in (6) above
but in the opposite direction (R).
(8) Disconnect power meter from RF OUTPUT. Connect 10 dB attenuator to RF
OUTPUT and connect power meter to attenuator.
(9) Press RECALL key and enter data 5 on keyboard. Add power meter indication
to exact value of attenuator (at 50 MHz) and record.
(10) Press RECALL key and enter data 6 on keyboard. Adjust A4R7 (fig. 11) for
power meter indication plus attenuator value (at 2.2 GHz) of equal difference from +10 dBm
as the value recorded in (9) above but in the opposite direction (R).
(11) Press RECALL key and enter data 2 on keyboard. Record power meter indication.
(12) Adjust FREQUENCY VERNIER control to tune TI output frequency to the
start of band 2 (approximately 2.5 GHz).