Power Supply
Figure 9. Test instrument boards A28 and A29.

Calibration Procedure For Oscilloscope Os-266(P)/U And Types 7844/R7844 Page Navigation
  14    15    16    17    18  19  20  
img
. . . . . . .
TB9-6625-2131-24
Table 6. Power Supply Check
Multimeter indications (V Dc)
Test instrument
test points
Min
Max
(A29 fig. 4)1
TP -15 SENSE
-14.85
-15.15
TP +5 SENSE
+4.9
+5.1
TP +15 SENSE
+14.85
+15.15
TP +50 SENSE
+49.5
+50.5
TP +130 SENSE
+124.8
+135.2
+5V LIGHTS CONTROL ILLUM.
+4.5
+5.5
Figure 9 for type R7844.
1
b. Adjustments
(1) Adjust -50 V ADJ A29 R1513 (fig. 4) (fig. 9 for type R7844) until multimeter
indicates 50 V dc (R).
(2) Adjust INVERTER CONTROL A28 R1293 (fig. 4) (fig. 9 for R7844) until
multimeter indicates between +45 and +55 V dc (R).
NOTE
Any change in A28 R1293 adjustment may affect the -50V dc
supply.
NOTE
Any change in A29 R1513 adjustment may affect the operation
of all TI circuits.