TB 9-6625-2132-24
(5) Press polarity pushbutton to + and adjust VOLTS counter to 0.100. Multimeter
will indicate between +0.0944 and +0.1051 V dc.
(6) Repeat technique of (5) above for VOLTS counter settings listed in table 4.
Multimeter will indicate within limits specified.
Test instrument
Multimeter indications
VOLTS counter settings
(V dc)
Min
Max
0.300
0.2947
0.3053
0.500
0.4945
0.5055
0.700
0.6943
0.7057
1.000
0.994
1.006
3.000
2.992
3.008
5.000
4.990
5.010
7.000
6.988
7.012
9.000
8.986
9.014
(7) Press polarity + (positive). Repeat technique of (1) and (2) above. If multimeter
does not indicate between +9.985 and +10.015 V dc, perform b (3) below.
(8) Press polarity - (negative) pushbutton. If multimeter does not indicate between -
9.985 and -10.015 V dc, perform b (4) below.
(9) Press polarity + (positive) pushbutton. Adjust COARSE and FINE controls
until VOLTS counter indicates 5.000. If multimeter does not indicate between +4.997 and
+5.003 V dc, perform b (5) below.
b. Adjustments
NOTE
Interaction exists between R573 and R575 (fig. 2).
If
adjustments were made, repeat a (1) through (5) above for best
compromise.
(1) Adjust R573 (fig. 2) for a 10 V dc indication on multimeter (R). Adjust for best
compromise between + and polarity.
(2) Adjust R575 (fig. 2) for a 1 V dc indication on multimeter (R). Adjust for best
compromise between + and - polarity.
(3) Adjust R686 (fig. 2) for a +10 V dc indication on multimeter (R).
(4) Adjust R727 (fig. 2) for a 10 V dc indication on multimeter (R).
(5) Adjust R696 (fig. 2) for a +5 V dc indication on multimeter (R).