TB 9-6625-2132-35
Test instrument VOLTS/DIV
Oscilloscope calibrator
Oscilloscope
switch settings
output settings (1 kHz)
vertical deflections (div)
.2 mV
1 mV
5
.5 mV
2 mV
4
2 mV
10 mV
5
5 mV
20 mV
4
10 mV
50 mV
5
20 mV
.1 V
5
50 mV
.2 V
4
.1 V
.5 V
5
.2 V
1 V
5
.5 V
2 V
4
1 V
5 V
5
2 V
10 V
5
5 V
20 V
4
10 V
10 V
5
16. Attenuation
Compensation
a. Performance Check
(1) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 output to TI
+INPUT, using termination.
(2) Set VOLTS/DIV switch to 50 mV and POSITION control to midrange.
(3) Press oscilloscope calibrator EDGE key and set oscilloscope calibrator output for
1 kHz and 5 divisions of vertical deflection on oscilloscope. If upper leading corner is not
square, adjust C241 (fig. 3) for optimum square wave (R).
(4) Remove connection from +INPUT and connect to -INPUT.
(5) Press -AC-GND-DC pushbutton to DC. If lower leading corner is not square,
adjust C141 (fig. 3) for optimum square wave (R).
(6) Set VOLTS/DIV switch to .1 V.
(7) Press +AC-GND-DC pushbutton to DC and -AC-GND-DC pushbutton to GND.
Figure 3. Adjustment locations -
left side view (Type 7A22).