TB 9-6625-2132-35
b. Adjustments
NOTE
Interaction exists between R573 and R575 (fig. 2).
If
adjustments were made, repeat a (1) through (5) above for best
compromise.
(1) Adjust R573 (fig. 2) for a 10 V dc indication on multimeter (R). Adjust for best
compromise between + and polarity.
(2) Adjust R575 (fig. 2) for a 1 V dc indication on multimeter (R). Adjust for best
compromise between + and - polarity.
(3) Adjust R686 (fig. 2) for a +10 V dc indication on multimeter (R).
(4) Adjust R727 (fig. 2) for a 10 V dc indication on multimeter (R).
(5) Adjust R696 (fig. 2) for a +5 V dc indication on multimeter (R).
Figure 2. Adjustment locations -
right side view (Type 7A13).
10. +INPUT and -INPUT Attenuator
Compensation
a. Performance Check
(1) Position TI controls as listed in (a) through (c) below:
(a) VOLTS/DIV switch to 10 mV.
(b) BW pushbutton to FULL.
(c) +INPUT pushbutton to DC.
(2) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 output to TI
+INPUT, using standardizer.
PIN: 057592-001