TB 9-6625-2132-35
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
Compatible with TI and supplied by
Tektronix, Type 7000 series
owner
Fluke, Model 5820A-5C-GHZ
Voltage output:
(5820A-5C-GHZ)
CALIBRATOR
Range: 1 mV to 100 V
Accuracy: 0.75%
Pulses:
Range: 100 Hz to 100 kHz
Risetime: 875 ps
Accuracy: 0.75%
TIME BASE
Compatible with TI and supplied by
Tektronix, Type 7B series
owner
TEKTRONIX, TYPE 7A13
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in the procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.