TB 9-6625-2132-35
e. Press time base pushbuttons as listed in (1) through (7) below:
(1) LEVEL/SLOPE pushbutton to POSITIVE SLOPE.
(2) MODE pushbutton to AUTO.
(3) COUPLING pushbutton to AC.
(4) SOURCE pushbutton to INT.
(5) MAGNIFIER pushbutton to X1 (IN).
(6) TIME/DIV or DLY TIME switch to 0.5 ms.
(7) TIME/DIV VARIABLE control to CAL.
f. Set oscilloscope ON/OFF switch to
ON and
allow at
least 20
minutes for warm-up.
8. Gain and
Deflection Accuracy
a. Performance Check
(1) Press TI +INPUT pushbutton to DC and set VOLTS/DIV switch to 1 mV.
(2) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to TI +INPUT.
(3) Press oscilloscope calibrator VOLTAGE key and set for a 5 mV, 1 kHz output.
(4) Turn oscilloscope calibrator EDIT FIELD knob to bring up oscilloscope
calibrator err display and set for 0.0 percent. Adjust TI GAIN (front panel) for 5 divisions
of vertical deflection on oscilloscope (R).
(5) Set TI and oscilloscope calibrator to values listed in table 3 and adjust
oscilloscope calibrator EDIT FIELD knob for 5 divisions of vertical deflection on
oscilloscope. Oscilloscope calibrator err display will be within 2 percent of output
b. Adjustments. No further adjustments can be made.
Test instrument
Oscilloscope calibrator
Oscilloscope
VOLTS/DIV
output settings (1 kHz)
vertical
switch settings
deflections (div)
2 mV
10 mV
5
5 mV
20 mV
4
10 mV
50 mV
5
20 mV
.1 V
5
50 mV
.2 V
4
.1 V
.5 V
5
.2 V
1 V
5
.5 V
2 V
4
1 V
5 V
5
2 V
10 V
5
5 V
20 V
4
CHANGE 2
PIN: 057592-002