Output Frequency

Calibration Procedure For Channel Alignment Indicator Id-1189/Pr Page Navigation
  5    6    7    8    9  10
img
. . .
TB 9-6625-2133-24
(9) Repeat technique of (4) and (6) through (8) above, for frequency settings listed in
table 3.
b. Adjustments. No further adjustments can be made.
21. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
10