(2) Set calibrator for a 10 mV dc output. If TI does not indicate between 9.995 and
10.005 mV, perform b below.
(3) Press TI pushbuttons and set calibrator output for settings listed in table 3. TI
will indicate within limits specified.
Test instrument indications
Test instrument
Calibrator output
pushbuttons
Min
Max
20 m
19
mV
18.992 mV
19.008
mV
200 m
190
mV
189.95
mV
190.05
mV
2
1.9
V
1.8995 V
1.9005
V
20
19
V
18.995 V
19.005
V
200
190
V
189.95
V
190.05
V
2000
990
V
989.7
V
990.3
V
b. Adjustments. Adjust GAIN ADJ R76 (fig. 1) for a TI indication of 10.000 (3
digits) (R).
Figure 1. Adjustment locations bottom view (models 3465A and 3465B).
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI V
and COM.
(2) Press TI
pushbutton.
(3) Press TI pushbuttons and set calibrator output for settings listed in table 4. At
each output setting, use calibrator output adjustment controls to set calibrator control
display Reading equal to TI indication. If calibrator control display ERROR indications
are not within limits specified, perform b below.