• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Equipment Setup
Ac Voltage

Calibration Procedure For Digital Multimeter Models 3465A, 3465B, And 3466A Page Navigation
  6    7    8    9    10  11  12    13    14  
img
. . . .
TB 9-6625-2135-24
b. Adjustments
(1) Set calibrator to STANDBY.
(2) Press TI RANGE 20 pushbutton.
(3) Set calibrator for a 19 V dc output. Adjust R602 19 VDC ADJ (fig. 2) for a 19.000
V TI indication (R).
Figure 2. Adjustment locations, model 3466A - top view.
17. Resistance
a. Performance Check
and COM.
(1) Connect calibrator OUTPUT HI and LO to TI INPUT V
and RANGE 20
pushbuttons.
(2) Press TI FUNCTION k
output and 2 wire Comp on. Adjust front panel
(3) Set calibrator for a 0
ZERO for a zero indication on TI.
12


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.