TB 9-6625-2135-35
Table 9. Ac Voltage - continued
Test instrument
Calibrator output
Test instrument indications
RANGE
pushbuttons
Voltage
Min
Max
2
1.9 V
5 kHz
1.8923
V
1.9077 V
2
1.9 V
15 kHz
1.877
V
1.923 V
2
1.9 V
100 kHz
1.847
V
1.953 V
20
19
V
20 Hz
18.57
V
19.43
V
20
19
V
40 Hz
18.78
V
19.22
V
20
19
V
5 kHz
18.923
V
19.077 V
20
19
V
15 kHz
18.77
V
19.23
V
20
19
V
100 kHz
18.47
V
19.53
V
200
190
V
20 Hz
185.7
V
194.3
V
200
190
V
40 Hz
187.8
V
192.2
V
200
190
V
5 kHz
189.23
V
190.77
V
200
190
V
15 kHz
187.7
V
192.3
V
200
100
V
100 kHz
96.5
V
103.5
V
1200 V
1000
V
40 Hz
987
V
1013
V
1200 V
1000
V
5 kHz
995
V
1005
V
1200 V
500
V
15 kHz
491
V
509
V
1200 V
100
V
100 kHz
83
V
117
V
b. Adjustments
(1) Set calibrator to STANDBY.
(2) Press TI RANGE 20 pushbutton.
(3) Set calibrator for a 19 V, 200 Hz output. Adjust R303 (fig. 2) for a 19.000 TI
indication (R).
(4) Set calibrator for a 1.9 V, 200 Hz output. Adjust R302 (fig. 2) for a 1.900 TI
indication (R).
(5) Repeat (3) and (4) above for best compromise.
(6) Set calibrator for a 19 V, 10 kHz output. Adjust R102 (fig. 2) for a 19.000 TI
indication (R).
(7) Set calibrator for a 1.9 V, 10 kHz output.
(8) Press TI RANGE 2 pushbutton. Adjust C102 (fig. 2) for a 1.9000 TI indication
(R).
(9) Set calibrator for a 1.9 V, 100 kHz output. Record TI indication.
(10) Press TI RANGE 20 pushbutton.
(11) Set calibrator for a 19 V, 100 kHz output. Adjust C202 (fig. 2) for a TI
indication numerically the same as recorded in (9) above (R).
(12) Set calibrator for a 1.9 V, 100 kHz output.
(13) Press TI RANGE 2 pushbutton and record TI indication.
18