TB 9-6625-2135-35
(3) Press TI range pushbuttons and set calibrator output for settings listed in
table 6. TI will
indicate within limits
specified.
Test instrument
Test instrument indications
range pushbuttons
Calibrator output
Min
Max
200
A
A
A
190
189.86
190.14
2m
1.9 mA
1.8986 mA
1.9014 mA
20m
19
mA
18.978 mA
19.022
mA
200m
190
mA
188.85
mA
191.5
mA
2000m
1.9 A
1888.5
mA
1911.5
mA
b. Adjustments. No adjustments can be made.
NOTE
Do not perform power supply check if all other parameters
are within tolerance.
a. Press TI function =V and 2 range pushbuttons.
b. Connect calibrator OUTPUT HI and LO to TI VΩ and COM.
c. Set calibrator for a 1 V dc output. If TI does not indicate 1.0000, adjust R89 (fig. 1)
for a 1.0000 TI indication (R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA Label/Form in accordance with TB 750-25.
SECTION IV
14. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 14 and
15 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.