TB 9-6625-2135-35
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
John Fluke, Model 5720A
Range: 10 mV to 1000 V
(p/o MIS-35947); w/power
Accuracy: (%)
amplifier, John Fluke,
Model 5725A (5725A)
Voltage
3565A & B
3566A
10 mV
.0125
---
19 mV
.0105
.0158
190 mV
.0066
.0132
1.9 to 190 V
.0066
.0092
500 V
---
.0150
990 V
.0076
---
1000 V
---
.0175
Ac voltage models 3465A and 3465B:
Range: 100 mV to 490 V
Frequency: 40 Hz to 20 kHz
Accuracy: (%)
Voltage
.04
.9
2
9
20
100 mV to
---
---
.04
.16
190 V
.04
490 V
.06
.06
.15
---
---
Ac voltage model 3466A
Range: 190 mV to 1000 V
Accuracy: (%)
Voltage
.02
.04
5
15
100
190 mV
to 190 V
.57
.29
.10
.10
.10
100 V
---
---
---
---
4.25
500 V
---
---
---
.45
---
1000 V
---
.32
.125
---
---
Dc current:
Range: 190 A to 1.9 A
Range: 100Ω to 10 M Ω
Accuracy: (%)
3465A & B
3466A
19Ω
---
.02275
190Ω
.00775
.02275
1.0 kΩ
.00775
---
1.9 to 190 kΩ
.00625
.00875
1.9 M Ω
.00625
.01125