TB 9-6625-2135-35
e. Short TI VΩ and COM. If necessary, adjust ZERO ADJ (rear panel) for 0.000
indication. Remove short.
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI VΩ and COM.
(2) Set calibrator for a 10 mV dc output. If TI does not indicate between 9.995 and
10.005 mV, perform b below.
(3) Press TI range pushbuttons and set calibrator output for settings listed in
table 3. TI will
indicate within limits
specified.
Test instrument indications
Test instrument
range pushbuttons
Calibrator output
Min
Max
20 m
19
mV
18.992
mV
19.008 mV
200 m
190
mV
189.95
mV
190.05
mV
2
1.9 V
1.8995
V
1.9005 V
20
19
V
18.995
V
19.005 V
200
190
V
189.95
V
190.05
V
2000
990
V
989.7
V
990.3
V
b. Adjustments. Adjust GAIN ADJ
R76 (fig. 1) for a TI
indication of
10.000 (3
digits) (R).
Figure 1. Adjustment locations (models 3465A and 3465B) -
bottom view.