TB 9-6625-2137-24
(5) Press TI MEASUREMENT/ENTRY
[OFFSET]
ENTRY
[RDNG
and
OFFSET] pushbuttons.
(6) Disconnect cable No. 1 from TI 75
input.
] pushbutton. TI will indicate less than or equal
(7) Press TERMINATION [600
to -40 dBmO.
(8) Reconnect cable No. 1 and repeat (2) through (7) above at 100 kHz.
(9) Remove balance test 600
adapter from setup.
NOTE
Steps (10) through (20) do not apply for model 3586C.
balance test - equipment setup.
NOTE
Cable No. 1 must be the same length and capacitance for 10 MHz
balance check.
(11) Repeat (2) through (6) above at 10 kHz.
10