Figure 5. Wideband power accuracy setup.

Calibration Procedure For Voltmeter, Frequency Selective An/Usm-490(), And Selective Level Meter, Models 3586B And 3586C W/Options 001, 003, And 004 Page Navigation
  11    12    13    14    15  16
TB 9-6625-2137-24
] pushbuttons.
(9) Press TI ENTRY [RECALL], [0], and TERMINATION [75
(10) Set synthesizer/level generator for a frequency of 50 kHz and an amplitude level
of -59.8 dBm.
(11) Press TI MEASUREMENT/ENTRY, FULL SCALE [ENTRY], and RANGE
[100 dB] pushbuttons.
(12) Press TI ENTRY [FULL SCALE], [3], [0], [MHz-dB], and [FREQ], [4], [8],
[9], [9] [6], [Hz MIN] pushbuttons.
(13) Press TI MEASUREMENT/MODE (blue) and [0 JITTER] pushbuttons.
Residual phase jitter will be less than or equal to 0.5 degrees p-p.
(14) Adjust synthesizer/level generator for 2 MHz.
Press ENTRY/[FREQ], [1],
[9],[9], [8], [9], [9], [6], and [Hz MIN] pushbuttons.
(15) The TI phase jitter reading will be less than or equal to 0.5 degrees p-p.
b. Adjustments. No further adjustments can be made.
19. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.
17/(18 Blank)