TB 9-6625-2138-35
Figure 6. Test instrument -
bottom interior view.
(2) Connect differential voltmeter between +15 V test point (fig. 6) and chassis
ground.
Differential voltmeter indication will be
between 14.5 and
15.75 V
dc.
NOTE
Top cover of TI must be removed to gain access to LOGIC
PROBE CONNECTOR (fig. 1).
(3) Connect differential voltmeter between center conductor of LOGIC PROBE
CONNECTOR and
chassis ground.
Differential voltmeter indication will be
between
4.75 and 5.25 V.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
15