TB 9-6625-2139-24
(40) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for 5
divisions of horizontal display. Oscilloscope calibrator err display will be within limits
Oscilloscope
Test instrument
Test instrument
Oscilloscope
calibrator
Test instrument
CH 1 VOLTS/DIV
A AND B SEC/DIV
calibrator VOLTAGE
Err display limits
adjustments
switch settings
output settings
switch settings
r%
X-Y
b(43) and (45)
10 m
50 mV at 1 kHz
3
b. Adjustments
(1) Position controls as listed in (a) through (c) below:
(a) HORIZONTAL MODE switch to A.
(b) A AND B SEC/DIV switches to .1 ms.
(c) X10 CAL control to in position.
(3) Adjust horizontal POSITION control to aline 1st time marker with the 1st
(extreme left) vertical graticule line.
(4) Adjust R740 A SWEEP GAIN (fig. 1) for 1 time marker per division over the
center 8 divisions. (R)
(5) Set HORIZONTAL MODE switch to B and adjust B INTENSITY control for
suitable viewing. Adjust horizontal POSITION control to aline 1st time marker with 1st
vertical graticule line.
(6) Adjust R730 B SWEEP GAIN (fig. 1) for 1 time marker per division over the
center 8 divisions. (R)
(7) Set HORIZONTAL MODE switch to A and pull X10 CAL control to out position.
(9) Adjust horizontal POSITION control to aline the nearest time marker to the 1st
vertical graticule line.
(10) Adjust R754 X10 GAIN (fig. 1) for 1 time marker per division. (R)
(11) Continue calibration at paragraph 9.a.(1) above.
(12) Set A AND B SEC/DIV switches to .2 ms.
(14) Adjust horizontal POSITION control to position middle time marker to center
vertical graticule line.
(15) Push X10 CAL control to in position.
23