TB 9-6625-2139-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Horizontal
Range: 0.5 s/div to 0.05 s/div
A sweep timing
Accuracy: 2%
Range: (X10 mag): 50 ms/div to 5 ns/div
Accuracy: 3%
B sweep timing
Range: 50 ms/div to 0.05 s/div
Accuracy: 2%
Range: (X10 mag): 5 ms/div to 5 ns/div
Accuracy: 3%
Sweep linearity
Accuracy: 5% (measured over any 2 of the center
8 divisions)
Deflection (X-Axis)
Range: 2 mV/div to 5 V/div
Accuracy: 3%
A trigger sensitivity
10
60
100
MHz
MHz
MHz
Internal
0.35
1.0
div
1.5
div
div1
External
35
120
150
mV
mV
mV2
B trigger sensitivity
Internal only
0.35 div
1.0
div
1.5
div
Calibrator amplitude
Range: 0.5 V
Accuracy: 2%3
10.3
division for type 2235.
2200 mV for type 2235.
3+5% for type 2235.
EQUIPMENT REQUIREMENTS
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI. Where the four-to-one ratio cannot be met, the four-to-one accuracy of the
equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in 4 above, and are not listed in this calibration
procedure.
The following peculiar accessory is also required for this calibration:
standardizer, 5-80 pF.