TB 9-6625-2139-35
Oscilloscope
calibrator
Oscilloscope
Err display
calibrator
Test instrument
Test instrument
indications
VOLTAGE
VOLTS/DIV
Test instrument
divisions of vertical
( %)
output settings
deflection
switch settings
adjustments
2m
10 mV
5
2
b(41) through (80)
5m
20 mV
4
2
10 m
50 mV
5
2
20 m
.1V
5
2
50 m
.2V
4
2
.1
.5V
5
2
.2
1V
5
2
.5
2V
4
2
1
5V
5
2
2
10 V
5
2
5
20 V
4
2
(13) Connect oscilloscope calibrator CHAN 1 to TI CH 1 using a 50Ω feedthrough
termination.
(14) Position controls as listed in (a) through (c) below:
(a) VERTICAL MODE CH 1 BOTH CH 2 switch to CH 1.
(b) CH 1 and CH 2 VOLTS/DIV switches to 2m.
(c) Set A AND B SEC/DIV switch to .05 s.
(15) Press oscilloscope calibrator EDGE pushbutton to illuminate green LED and set
oscilloscope calibrator output to 10 mV at 1 MHz.
(16) Use technique of step 17 below for TI settings and oscilloscope calibrator output
(17) Adjust CH 1 POSITION control to position top of waveform to center horizontal
graticule line. If squarewave aberrations exceed those listed in table 5, perform