TB 9-6625-2151-35
(3) Connect multimeter Input LO to JU-14 (fig. 4) and HI to ground. Multimeter
will indicate between -13.75 and -14.25 V.
b. Adjustments
(1) Adjust R42 +5V ADJ (fig. 4) for a +5 V dc indication on multimeter.
(2) Adjust R31 +14V ADJ (fig. 4) for a +14 V dc indication on multimeter.
Figure 4. Transmitter A14 board -
component locations.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA Label/Form in accordance with TB 750-25.