• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 14. Winslow, Models 334W and 334X
Table 15. General Radio, Model 1434 Series

Calibration Procedure For Resistance Decades And Resistance Standards (General) Page Navigation
  20    21    22    23    24  25  26    27    28    29  
img
TB 9-6625-2153-24
Table 14. Winslow, Models 334W and 334X - Continued
Test instrument
Resistance indications
Decade switch
Min
Max
Decade switch
positions
Ohms
1.0:/step (cont)
6
05.98900
06.01100
7
06.98800
07.01200
8
07.98700
08.01300
9
08.98600
09.01400
10
09.98500
10.01500
1
09.98500
10.01500
10:/step
2
19.97500
20.02500
3
29.96500
30.03500
4
39.95500
40.04500
5
49.94500
50.05500
6
59.93500
60.06500
7
69.92500
70.07500
8
79.91500
80.08500
9
89.90500
90.09500
10
99.89500
100.10500
1
99.89500
100.10500
100:/step
2
199.79500
200.20500
3
299.69500
300.30500
4
399.59500
400.40500
5
499.49500
500.50500
6
599.39500
600.60500
7
699.29500
700.70500
8
799.19500
800.80500
9
899.09500
900.90500
10
998.99500
1001.00500
Kilohms
1
0.99900000
1.0010000
1 k:/step
2
1.9980000
2.0020000
3
2.9970000
3.0030000
4
3.9960000
4.0040000
5
4.9950000
5.0050000
6
5.9940000
6.0060000
7
6.9930000
7.0070000
8
7.9920000
8.0080000
9
8.991000
9.009000
10
9.990000
10.010000
1
9.990000
10.010000
10 k:/step
2
19.980000
20.020000
3
29.970000
30.030000
4
39.960000
40.040000
5
49.950000
50.050000
6
59.940000
60.060000
7
69.930000
70.070000
8
79.920000
80.080000
9
89.910000
90.090000
10
99.900000
100.100000
1
This decade applicable to model 334W only.
26


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.