• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 12. General Resistance, Model RDS-615B
Table 14. Winslow, Models 334W and 334X

Calibration Procedure For Resistance Decades And Resistance Standards (General) Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
img
. . . .
TB 9-6625-2153-24
Table 12. General Resistance, Model RDS-615B - Continued
Test instrument
Resistance indications
Decade switch
Min
Max
Decade switch
positions
Kilohms
9
8.9995500
9.0004500
1 k:/step (cont)
10
9.9995000
10.000500
1
9.9995000
10.000500
10 k:/step
2
19.999000
20.001000
3
29.998500
30.001500
4
39.998000
40.002000
5
49.997500
50.002500
6
59.997000
60.003000
7
69.996500
70.003500
8
79.996000
80.004000
9
89.995500
90.004500
10
99.995000
100.00500
1
99.995000
100.00500
100 k:/step
2
199.99000
200.01000
3
299.98500
300.01500
4
399.98000
400.02000
5
499.97500
500.02500
6
599.97000
600.03000
7
699.96500
700.03500
8
799.96000
800.04000
9
899.95500
900.04500
10
999.95000
1000.05000
Table 13. ICC, Models CR10M, and CR100M, and CR1000M1
Test instrument
Performance specifications
CR10M
Range: 10 M: in 1 M: steps
Accuracy: r0.02%
CR100M
Range: 100 M: in 10 M: steps
Accuracy: r0.05%
CR1000M
Range: 1000 M: in 100 M: steps
Accuracy: r0.25%
Resistance indications
Min
Max
CR10M
999.80000
k:
01.000200
M:
CR100M
09.995000  M:
010.00500
M:
CR1000M
099.75000
M:
100.2500
M:
1Repeat measurement for each pair of resistance terminals (2 and 3, 3 and 4, 4 and 5, etc.).
Resistance measuring system
will indicate within limits specified for TI model as listed in table 13. Prepare a calibration test report for all TIs calibrated at
reference level if the TI is not within limits specified in table 13. When calibrating at transfer level, if the TI is not within limits
specified in table 13, the TI will be evacuated to the ACL for calibration support. Once a test report is prepared, all subsequent
support for the TI will be at the ACL. When TI is repaired, and measures within specified limits, a test report is no longer
required and item will revert to original support level (reference or transfer).