Table 15. General Radio, Model 1434 Series
Table 16. Leeds and Northrup, Models 4755 and 4776

Calibration Procedure For Resistance Decades And Resistance Standards (General) Page Navigation
  22    23    24    25    26  27  28    29  
img
TB 9-6625-2153-24
Table 15. General Radio, Model 1434 Series - Continued
Test instrument
Resistance indications
Decade switch
Decade switch
Min
Max
positions
Ohms
6
599.88000
600.12000
100:/step (cont)
7
699.86000
700.14000
8
799.84000
800.16000
9
899.92000
900.18000
10
999.80000
1000.20000
0
---
---
Kilohms
1
0.9998000
1.0002000
1 k:/step
2
1.9996000
2.0004000
3
2.9994000
3.0006000
4
3.9992000
4.0008000
5
4.9990000
5.0010000
6
5.9988000
6.0012000
7
6.9986000
7.0014000
8
7.9984000
8.0016000
9
8.9982000
9.0018000
10
9.9980000
10.0020000
0
---
---
1
9.9980000
10.0020000
10 k:/step
2
19.996000
20.004000
3
29.994000
30.006000
4
39.992000
40.008000
5
49.990000
50.010000
6
59.988000
60.012000
7
69.986000
70.014000
8
79.984000
80.016000
9
89.982000
90.018000
10
99.980000
100.020000
0
---
---
1
99.980000
100.020000
100 k:/step
2
199.96000
200.04000
3
299.94000
300.06000
4
399.92000
400.08000
5
499.90000
500.10000
6
599.88000
600.12000
7
699.86000
700.14000
8
799.84000
800.16000
9
899.82000
900.18000
10
999.80000
1000.20000
0
---
---
1Use
appropriate ranges for particular model of TI being tested.
Accuracy for 100 : Rheostat (1434 QC only) is r 1 :