TB 9-6625-2153-24
Table 15. General Radio, Model 1434 Series - Continued
Test instrument
Decade switch
Decade switch
Min
Max
positions
Ohms
6
599.88000
600.12000
100:/step (cont)
7
699.86000
700.14000
8
799.84000
800.16000
9
899.92000
900.18000
10
999.80000
1000.20000
0
---
---
Kilohms
1
0.9998000
1.0002000
1 k:/step
2
1.9996000
2.0004000
3
2.9994000
3.0006000
4
3.9992000
4.0008000
5
4.9990000
5.0010000
6
5.9988000
6.0012000
7
6.9986000
7.0014000
8
7.9984000
8.0016000
9
8.9982000
9.0018000
10
9.9980000
10.0020000
0
---
---
1
9.9980000
10.0020000
10 k:/step
2
19.996000
20.004000
3
29.994000
30.006000
4
39.992000
40.008000
5
49.990000
50.010000
6
59.988000
60.012000
7
69.986000
70.014000
8
79.984000
80.016000
9
89.982000
90.018000
10
99.980000
100.020000
0
---
---
1
99.980000
100.020000
100 k:/step
2
199.96000
200.04000
3
299.94000
300.06000
4
399.92000
400.08000
5
499.90000
500.10000
6
599.88000
600.12000
7
699.86000
700.14000
8
799.84000
800.16000
9
899.82000
900.18000
10
999.80000
1000.20000
0
---
---
1Use
appropriate ranges for particular model of TI being tested.
Accuracy for 100 : Rheostat (1434 QC only) is r 1 :
2
28