TB 9-6625-2163-35
C4
B1
D1
B2
D2
B4
D4
F2
I/P
A2
A4
C1
A1
C2
A11TP1
A11TP2
(8) Repeat technique of (7) above for each of the following pulses: A1, C2, A2, C4,
A4, B1, D1, B2, D2, B4, D4, F2, and I/P (fig. 3). If pulse positions are not as specified,
perform b(2) below for appropriate pulses.
b. Adjustments
(1) Adjust oscilloscope HORIZONTAL POSITION control to reference C1 pulse on
CH 1 and
CH 2 to
center graticule line on
display. Adjust TI
control A7R1 (fig. 2) to
center
(2) Adjust HORIZONTAL POSITION control to reference respective pulse on CH
1 and CH 2 to center graticule line on display. Adjust controls as indicated in (a) through
(m) below (fig. 3) to center pulses on CH 2 within respective pulse on CH 1 (fig. 3):
(a) Pulse A1
A7R2 (R)
(b) Pulse C2
A7R3 (R)
(c) Pulse A2
A7R4 (R)
(d) Pulse C4
A7R5 (R)
(e) Pulse A4
A7R6 (R)
(f) Pulse B1
A7R7 (R)
(g) Pulse D1
A7R8 (R)
(h) Pulse B2
A7R9 (R)
(i) Pulse D2
A7R10 (R)
(j) Pulse B4
A7R11 (R)
(k) Pulse D4
A7R12 (R)
(l) Pulse F2
A7R13 (R)
(m) Pulse I/P
A7R14 (R)
12