TB 9-6625-2164-35
az. Press TRIGGERING/MODE P/P pushbutton.
ba. Adjust TRIGGERING LEVEL control fully cw.
bb.Adjust CURSORS ⇐ 1 ⇒ T=0.0S. Triggering point (CURSOR dot) will indicate
between 0.8 and 1.2 divisions below sine wave positive peak; if not, adjust P-P RANGE
R541 (fig. 3) for the (CURSOR dot) position to fall between 0.8 and 1.2 divisions below sine
wave positive peak.
bc. Adjust TRIGGERING LEVEL control fully ccw, and press - SLOPE (+SLOPE
pushbutton not lighted).
bd.Triggering point (CURSOR dot) will indicate between 0.8 and 1.2 divisions above
sine wave negative peak; if not, adjust P-P RANGE R541 (fig. 3) for best compromise to
meet requirement of bb through bd above.
be. Press TRIGGERING/MODE AUTO and TRIGGERING/+SLOPE pushbuttons.
bf.
Connect multimeter between TP400 and GND REF
TP900 (fig. 2).
bg. Adjust TRIGGERING LEVEL control for -1.28 V, + 10 mV indication on multimeter.
bh.
Set CH 1 VOLTS/DIV switch to
1 V, and
set function/arbitrary generator output
frequency to 1 kHz and amplitude for 0.4 divisions on TI display.
bi. The 1 kHz sine wave will remain triggered while switching between AUTO +
SLOPE and AUTO - SLOPE without readjusting TRIGGER LEVEL control; if not,
repeat steps w through au above.
bj. While switching between the P-P + SLOPE and P-P - SLOPE, observe that the
1 kHz remains triggered without adjusting the TRIGGER LEVEL.
NOTE
If the requirements of bi and bj above are met, ignore the
remainder of this step (bk through bs) below.
bk.Press CH 1 GND pushbutton.
bl. Press TRIGGERING MODE/AUTO, COUPLING/AC and + SLOPE pushbuttons.
bm.
Connect multimeter between TP400 and GND REF
TP900 (fig. 2).
bn.
Adjust TRIGGERING LEVEL control for -1.28 V, + 10 mV
indication on
multimeter .
bo. Move the
positive lead of
multimeter to
TP540 and
adjust PLUS PEAK
OFFSET
R546 (fig. 3) for + 10.0 mV indication on multimeter.
9