TB 9-6625-2170-24
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-287 or AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories issued as indicated in 4 above, and are not listed in this calibration
procedure.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range: 3 mV to 100 V
Fluke, Model 5720A (5720A)
(p/o MIS-35947)
1 r0.5%
Accuracy:
Fluke, Model 8840A/AF05
Range: 4.98 to 5.51 V dc
(AN/GSM-64D)
Accuracy: r0.476%
Capabilities
Agilent, OS-3-3/G (OS-303/G)
Vertical: 200 mV/div dc coupling
Horizontal: 200 ms/div
Range: 0.001
ESI, Model DT72A (7915908)
Frequency: 1 kHz
Accuracy:1
Tektronix, Type SG-505
Amplitude range: -10 to 0 dBm
(MIS-30526/11)
Distortion: <0.005%
Range: 1 mV to 1.58 V
Fluke, Model 8922A/AA
(8922A/AA)
Accuracy: r0.75%
1Combined accuracy of calibrator and ratio transformer for 100 PV, 1 kHz output is r0.5%.
CALIBRATION PROCESS
6. Preliminary Instructions
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common