TB 9-6625-2171-35
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287. Alternate items may be used by the calibrating activity when
the equipment listed in table 2 is not available. The items selected must be verified to
perform satisfactorily prior to use and must bear evidence of current calibration. The
equipment must meet or exceed the minimum use specifications listed in table 2. The
accuracies listed in table 2 provide a four-to-one ratio between the standard and TI
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Termination, N Jack, dc to 5 GHz, 50 W, Sierra, Model 161A-50 (161A-50) and
Low Pass Filter, Telonic, Models TLC125-6EF1 and TLC75-6EF1.
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
AUDIO ANALYZER
Distortion capability: 3 to 15%
Boonton, Model 1120-S/10
(MIS-35954/2)
Hewlett-Packard, Model 3458A/E02
Range: 2 mV to 10 V ac at
(MIS35947/1)
30 to 1020 Hz;
5.5 V dc
Accuracy: 1%
DUAL VOLT POWER
Range: -8 to +8 V dc
Tektronix, Type PS503A (MIS-30526/6)
Accuracy: 1%
SUPPLY
Range: 29 Hz to 334.7 MHz
Fluke, Model PM6681
Accuracy: 0.000625%
MEASURING RECEIVER
Range: 5 to 44% modulation
Hewlett-Packard, Model 8902A
Accuracy: 2.5%
with power sensor Hewlett-Packard,
Model 11722A (7917002 or 7911261)
Sweep range: 0.1 ms to 10 s
Tektronix, Model 2465B (OS288/G)
Vertical range: 0.5 V/div
Accuracy: 3%
Hewlett-Packard, Model E12-432A
POWER METER
Range: -10 to -20 dBm
Accuracy: .5%
(MIS-30525) w/thermistor mount,
Hewlett-Packard, Model 478A-H75
Range: 1 to 10 mW
Accuracy: 3%
POWER STANDARD
Range: 50 MHz at 10 to 40 W
Directional Coupler, Maury, Model
ASSEMBLY
4098A (7916259)
RF POWER AMPLIFIER
Range: 30 to 80 MHz at
3 to 40 W
(MIS45845)
Range: 30 to 80 MHz
Wiltron, Model 68347M
Accuracy: 2 kHz
SPECTRUM ANALYZER
Range: 30 to 80 MHz
(AN/USM-489A)
Sensitivity: -101 dBm
Range: 0 to 360
ZIFOR
Collins, Model 478A-3 (478A-3)
Accuracy: 0.02