TB 9-6625-2182-24
Measuring receiver
Test instrument
indications
DATA
(dBm)
amplitude
Min
Max
15 dBm
13
17
10 dBm
8
12
5 dBm
3
7
0 dBm1
2
-2
-10 dBm
-8
-12
-20 dBm
-18
-22
-30 dBm
-28
-32
-40 dBm
-38
-42
-50 dBm
-48
-52
-60 dBm
-58
-62
-70
dBm
-68
-72
-80
dBm
-78
-82
-90
dBm
-88
-92
-100
dBm
-98
-102
-110
dBm
-108
-112
1Setup
measuring receiver for tuned RF level cal techniques and
wait for receiver to calibrate.
(6) Press RF OFF/ON pushbutton to OFF.
b. Adjustments. No adjustments can be made.
11. Output Level Flatness
a. Performance Check
(1) Set measuring receiver with sensor module to measure RF power with
0.01 db resolution.
(2) Press pushbuttons ENTRY AMPTD and DATA (+10 dBm).
(3) Measure and record the RF power using the measuring receiver, while using the
ENTRY FREQ key and the DATA key pad to select frequencies between the start and
(4) Calculate the flatness using the formula below. The flatness will be less than or
equal to the maximum limits listed in table 7.
Flatness = (highest lowest)/2