TB 9-6625-2185-35
(15) Adjust R44 (fig. 1) until TI displays 3685 (R).
(16) Repeat (11) above.
(17) Repeat 10a above.
11. Sensor Frequency
Response
a. Performance Check
(2) Replace thermistor mount H75-478A with thermistor mount 8478B and allow
warm-up of power meter.
(3) Connect equipment as shown in figure 6, CONNECTION A.
Figure 6. Frequency response -
equipment setup.
(4) Adjust signal generator for a 1 GHz, 1 mW output as indicated on power
meter.
(5) Calculate power using precision power measurement technique listed in
(6) Press MODE PWR, X (where X is the frequency being measured) and CAL
FAC GHz pushbuttons.
(7) Connect equipment as shown in figure 6, CONNECTION B. If display does
not indicate within 3 percent of power calculated in (5) above, perform b below.
23