TB 9-6625-2193-35
EQUIPMENT REQUIREMENTS
4. Equipment Required.
Table 2 identifies the
specific equipment to be
used in
this calibration procedure. This equipment is issued with Secondary Transfer
Calibration Standards Set AN/GSM-286. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must
meet or exceed the minimum use specifications listed in table 2. The accuracies
listed in table 2 provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The
accessories required for this
calibration are
common usage accessories issued as indicated in paragraph 4 above and are not
listed in this calibration procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Ac voltage:
John Fluke, Model
Range: 3.16 mV to 300 V
5720A(5700A/EP) (p/o MIS-35947);
Frequency: 20 Hz to 1 MHz
w/power amplifier John Fluke,
Accuracy: (%)
5725A) (5725A)
Frequency:
20 Hz, 100 kHz & 500 kHz...1.25
1 and 50 kHz......................0.75
1 MHz...............................2.50
Wideband voltage:
Voltage: 316 V to 3.0 V
(1 kHz reference)
Amplitude flatness: (%)
5 MHz................2.50
20
MHz...............3.75
Range: 4.9 to 15 V dc
John Fluke, Model 8840A/AF-05/09
Accuracy: 3.3 %
(AN/GSM-64D)
ESI DT72A
Range: 0.001
(7915908)
Frequency:50 Hz, 1 kHz
Accuracy:1
Combined accuracy of calibrator and ratio transformer for 100 V and 316.2 V at 50 Hz and 1 kHz output is 0.75%.
1