TB 9-6625-2193-35
Figure 2. View of
underside of
module assembly.
(4) Connect TI INPUT to calibrator WIDEBAND output and press
calibrator W BND pushbutton.
(5) Set calibrator for an initial 316 V, 1 kHz wideband output. Adjust
calibrator for a TI indication equal to value recorded in table 3 for 316 V (1 kHz) to
establish a 1 kHz reference. Press calibrator NEW REF pushbutton.
(6) Set calibrator frequency to 500 kHz, then readjust amplitude for TI
reference established in (5) above. If calibrator Error display readout does not
indicate within 5%, perform b below.
(7) Repeat technique of (6) above for remaining frequencies listed for the
316.2 V range in table 4. Calibrator Error display indication will be within the
limits specified.
(8) Repeat technique of (5) through (7) above for remaining calibrator initial
voltage and frequencies listed in table 4.
8 CHANGE 3