TB 9-6625-2193-35
b. Move
positive lead to TP2
(fig. 1). If
digital multimeter
does not
indicate -
15.0 (0.5) V, adjust R49 (fig. 3) for a -15.0 V indication (R).
c. Move
positive lead to TP3
(fig. 3). Digital multimeter will
indicate +
5.0
(0.25) V.
d. Move
positive lead to pin 1 of
R34 (fig. 3). Digital multimeter will
indicate
+9.3 (1.0) V.
e. Move
positive lead to
TP10 (fig. 3). Digital multimeter will
indicate +4.9
(0.4) V.
Figure 3. View of
underside of
module assembly.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.