TB 9-6625-2197-40
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Reference Standards Set
NSN 4931-00-621-7878 and limited deployed to Secondary Transfer Calibration Standards
Set, AN/GSM-287 or AN/GSM-440. Alternate items may be used by the calibrating activity.
The item selected must be verified to perform satisfactorily prior to use and must bear
evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI. Where the four-to-one ratio cannot be met, the four-to-one
accuracy will be listed, and the actual accuracy of the equipment selected is shown in
parentheses.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Transponder test set kit, Teledyne Electronics, P/N 154397 (NSN 6625-01-118-
4440), battery charger assembly (45413-156721), and Circulator (791684).
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
FREQUENCY COUNTER
Frequency: 1030 MHz
Fluke, Model PM6681/656
Accuracy: 0.005 %
(PM6681/656)
Range: -12 to +12 V dc
Agilent, Model 3458A (3458A)
Accuracy: 0.02%
POWER METER
Wavetek, Model 8502A (8502-16934-
Power range: + 6 to - 4 dBm
17071) w/power detector, Wavetek,
Accuracy: 0.5 dBm
Model 16934 (16934)
Agilent, Model 81150A Pulse
Repetition rate: 10 to 50 kHz
Pulse width: 0.1 to 1 sec
(81150A)
Amplitude: 1 to 5 volts
RADAR TEST SET 1
(AN/UPM155)
Power output : 0 to - 30 dBm
Accuracy: 0.5 dBm (1.0 dBm)
1 Non
set configuration.
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instruction outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.