TB 9-6625-2204-35
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories listed in
table 3 are locally procured items
which are used in this calibration procedure. Interconnecting leads and adapters used in
this procedure are issued with Secondary Calibration Standards Set AN/GSM-286.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Fluke, Model 8840A/AF05 (AN/GSM-64D)
Range: 1 to 30.2 V dc
106 to 124 V ac
Accuracy: 1%
Range: 3Ω to 120Ω
Accuracy: 0.5%
Fluke, Model 5720A (5700A/EP) (p/o MIS-
CALIBRATOR
Range: 9.6 to 31.2 V dc
35947); w amplifier, Fluke 5725A/AR
4.6 to 10.5 mA dc
(5725A/AR)
9.5 to 10.5 V ac
Accuracy: 1%
Range: 4.5 to 5.5 V
(OS 303/G)
0.1 to 5.5 s
Accuracy: 3%
Range:
LeCroy, Model 9210 (9210) w/plug-ins, LeCroy,
Models 9211 (9211) and 9215 (9215)
(MIS 45839)
0.5 s
Width:
Amplitude: 5.0 V
Common name
Description (part number)
RESISTOR
RESISTOR1
RESISTOR2
RESISTOR
RESISTOR
RESISTOR
1Three
required.
2Four
required.