TB 9-6625-2204-35
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 11-6625-842-15 for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Mechanically zero TI panel meters if required.
b. Set 6 AMP POWER circuit breaker to OFF and connect TI to 115 V ac, 400 Hz,
using cable W3 (P/0 TI).
c. Construct test loads as
shown in
figure 1A and B.
d. Using multimeter, measure resistance value of ac
load (fig. 1A). If
resistance value
exceeds 117.8 Ω, add 4700 Ω, resistor as shown in figure 1A.
e. Repeat d
above for dc
load (fig. 1B). If
resistance value exceeds 4.l Ω,
add 160 Ω
resistor as shown in figure 1B. If dc load is less than 3.9 Ω, add 0.1 Ω resistor as shown in
appropriate lead. Connect other end of ac load to TRANSPONDER J2 pins C14 and C13