• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Dc Voltage
Dc Current

Calibration Procedure For Digital Multimeter, Mis-28709D (Model 310) And Model 6V7070 Page Navigation
  1    2    3  4  5    6  
img
. . . . . . .
TB 9-6625-2208-24
Figure 1. Adjustment location.
9. Ac Voltage
a. Performance Check
(1) Set the function/ range switch to ACV 200 m position.
(2) Set calibrator for an output amplitude of 190 mV and an output frequency of 45 Hz.
TI will indicate within limits specified in first row of table 4.
(3) Repeat technique of (1) and (2) above using settings listed in table 4. TI will
indicate within limits specified in table 4.
Table 4. Ac Voltage
Test instrument1
Calibrator
Output
Output
Range
Indication limits
amplitude
frequency
setting (ACV)
Min
Max
190.0 mV
45 Hz
200 m
188.3
191.7
190.0 mV
1 kHz
200 m
188.3
191.7
1.9 V
45 Hz
2
1.883
1.917
1.9 V
1 kHz
2
1.883
1.917
1.9 V
3 kHz
2
1.867 (1.883)
1.934 (1.917)
1.9 V
5 kHz
2
1.844 (1.883)
1.957 (1.917)
19.0 V
1 kHz
20
18.83
19.17
190.0 V
1 kHz
200
188.3
191.7
1000  V
1 kHz
1000
990 (980)
1011 (1020)
Values in parenthesis are for model 6V-7070.
1
(4) Reduce all outputs to minimum.
b. Adjustments. No adjustments can be made.
5


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.