TB 9-6625-2208-24
Table 6. Resistance
Test instrument
Calibrator
err
display indication
Function/range
(%)
switch settings
Nominal output
190
0.576
200
1.9 k
0.576
2 k
19  k
0.576
20 k
190  k
0.576
200 k
1.9 M
0.576
2 M
19  M
1.53
20 M
b. Adjustments. No adjustments can be made.
12. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
7/(8 BLANK)