TB 9-6625-2215-35
b. Mount a calibrated TD-1209/U (Hewlett-Packard, Model 5300A) measuring system
on TI and orient the interconnected units upside down.
c. Connect measuring system to
autotransformer and connect autotransformer to a
115 V
ac source.
e. Adjust measuring system SAMPLE RATE control cw to ON position and allow 1
hour for warm-up.
f. Set INT STD-EXT STD switch on
rear panel of
frequency counter to
EXT STD.
NOTE
Some models do not have an internal oscillator. When calibrating
these models, omit this check.
a. Performance Check
(1) Connect frequency counter A input to TI OSC jack (rear panel).
of exact frequency stamped on TCXO case (fig. 1).
Figure 1. Test instrument -
(Hewlett-Packard, Model 5303B) -
bottom view.