TB 9-6625-2215-35
Test instrument
Measuring
RANGE pushbutton settings
system
(V)
indications
0.0000 (0.0020)
10
100
00.000 (00.050)
000.00 (000.50)
1000
Figure 3. Test instrument -
(Hewlett-Packard, Model 5306A) -
bottom view.
(4) Connect calibrator to TI HI and LO terminals.
(5) Press RANGE 10 V pushbutton.
(6) Set calibrator output to 9.000 V at 40 Hz. If TI does not indicate between 8.9098
and 9.0902 V ac, perform b (2) below.
(7) Repeat technique of (5) and (6) above, using values listed in table 4.